Have done the measurement output plane x-ray radiation dose estimation for patients in the examination of the thorax, abdomen and skull. Parameters measured (1) Measurement accuracy voltage tubes, (2) the accuracy of the Measurement of time, (3) eksposi Linieritas output or the rate of radiation dose, (4) the stability of the voltage of the tube, eksposi and time radiation output, (5) the quality of the x-ray beam. The research of using conventional x-ray unit Toshiba brand model DRX-1824B, and the unfors xi detector sets. The research results indicate (1) aircraft of conventional x-ray unit Toshiba brand model DRX-1824B are in good condition, (2) the average value of estimated doses of patients in the examination of the thorax, abdomen and the skull does not exceed the dose limit value permitted by BAPETEN.Keywords: radiation dose, the abdomen, the aircraft of a conventional x-ray, skull, thorax